xraytools.

Intensity Corrections

A diffractometer records counts; a structure needs |F|². Between them sit the Lorentz and polarisation factors, which are pure geometry and optics, and absorption, which for the flat plate almost everyone owns turns out to be a constant. This page takes the factor apart, draws it, and runs the whole chain backwards on a structure whose answer is already known.

You supply
Nothing of your own. Pick one of the structures this site ships, a wavelength and a monochromator setting; the intensities come from the simulator next door, so the answer is known before the correction is applied. Sample thickness is used only by the absorption panel.
Reading it
The strongest peak in a pattern is usually not the strongest reflection, and Lp is most of the difference: it spans two orders of magnitude across a scan, so the ranking you read off a chart is not the ranking of |F|². Lp also does not simply fall with angle — it turns round near 2θ = 98° and climbs again. The one correction people leave out is the monochromator, and leaving it out is not a scale error: the residual grows with angle, which is where an overall displacement parameter lives.

Worked examples: quartz – a reflection lying fifteen places from where the chart puts it · rutile with no monochromator – a heavy absorber, ten micrometres deep · quartz at Cr Kα – the depth panel refuses, and says why

See also: Absorption Coefficient Calculator · HKL Calculator · Peak Finding · Wilson Plot and E Statistics · Structure Factor Calculator

Input

The measurement

This wavelength is Cu Kα.

The instrument
°

The angle is used only when the setting above is “give the monochromator angle myself”. For graphite it is computed from the wavelength, so changing the anode moves it.

mm

What a measured intensity actually is

A diffractometer records counts. What a structure determination needs is |F|², and everything between the two is geometry and optics:

Iobs = s · L(θ) · P(θ) · A(θ) · m · |F

s is one scale factor for the whole data set and carries the beam current, the counting time and the sample volume; nobody knows it and nobody needs to, because a structure is refined against relative intensities. The other four are the ones that vary from line to line, and those cannot be scaled away:

  • L, the Lorentz factor — how long a reflection spends in the diffracting condition, plus, for a powder, what fraction of its Debye ring the detector sees. Pure geometry, no physics of the sample in it at all.
  • P, the polarisation factor — an X-ray scattered through 2θ keeps only the component of its field perpendicular to the scattering plane, so scattering near 90° is suppressed. A monochromator polarises the incident beam before the sample ever sees it, which changes this factor and is the correction people forget.
  • A, absorption — how much of the beam survives the trip in and out. For the flat plate almost everyone owns this turns out to be a constant, and the panel below shows why that is a fact about the geometry rather than about the sample.
  • m, multiplicity — a powder cannot separate reflections that land at the same angle, so the peak carries their sum. The simulator counts them.

This page takes L and P apart, shows what A does, and then runs the whole chain backwards on a structure whose answer is known.

The factor, drawn

0.1110100100030609012015098.4°2θ 5° — L 131.5208, P 0.9962, Lp 131.02132θ 5° — L 131.5208, P 0.9962, Lp 131.02132θ 10° — L 33.0372, P 0.9849, Lp 32.53912θ 10° — L 33.0372, P 0.9849, Lp 32.53912θ 15° — L 14.8005, P 0.9665, Lp 14.30482θ 15° — L 14.8005, P 0.9665, Lp 14.30482θ 20° — L 8.4188, P 0.9415, Lp 7.92642θ 20° — L 8.4188, P 0.9415, Lp 7.92642θ 25° — L 5.4662, P 0.9107, Lp 4.97802θ 25° — L 5.4662, P 0.9107, Lp 4.97802θ 30° — L 3.8637, P 0.8750, Lp 3.38072θ 30° — L 3.8637, P 0.8750, Lp 3.38072θ 35° — L 2.8989, P 0.8355, Lp 2.42212θ 35° — L 2.8989, P 0.8355, Lp 2.42212θ 40° — L 2.2743, P 0.7934, Lp 1.80452θ 40° — L 2.2743, P 0.7934, Lp 1.80452θ 45° — L 1.8478, P 0.7500, Lp 1.38582θ 45° — L 1.8478, P 0.7500, Lp 1.38582θ 50° — L 1.5444, P 0.7066, Lp 1.09132θ 50° — L 1.5444, P 0.7066, Lp 1.09132θ 55° — L 1.3219, P 0.6645, Lp 0.87842θ 55° — L 1.3219, P 0.6645, Lp 0.87842θ 60° — L 1.1547, P 0.6250, Lp 0.72172θ 60° — L 1.1547, P 0.6250, Lp 0.72172θ 65° — L 1.0268, P 0.5893, Lp 0.60512θ 65° — L 1.0268, P 0.5893, Lp 0.60512θ 70° — L 0.9277, P 0.5585, Lp 0.51812θ 70° — L 0.9277, P 0.5585, Lp 0.51812θ 75° — L 0.8503, P 0.5335, Lp 0.45362θ 75° — L 0.8503, P 0.5335, Lp 0.45362θ 80° — L 0.7899, P 0.5151, Lp 0.40682θ 80° — L 0.7899, P 0.5151, Lp 0.40682θ 85° — L 0.7429, P 0.5038, Lp 0.37432θ 85° — L 0.7429, P 0.5038, Lp 0.37432θ 90° — L 0.7071, P 0.5000, Lp 0.35362θ 90° — L 0.7071, P 0.5000, Lp 0.35362θ 95° — L 0.6808, P 0.5038, Lp 0.34302θ 95° — L 0.6808, P 0.5038, Lp 0.34302θ 100° — L 0.6628, P 0.5151, Lp 0.34142θ 100° — L 0.6628, P 0.5151, Lp 0.34142θ 105° — L 0.6525, P 0.5335, Lp 0.34812θ 105° — L 0.6525, P 0.5335, Lp 0.34812θ 110° — L 0.6496, P 0.5585, Lp 0.36282θ 110° — L 0.6496, P 0.5585, Lp 0.36282θ 115° — L 0.6541, P 0.5893, Lp 0.38552θ 115° — L 0.6541, P 0.5893, Lp 0.38552θ 120° — L 0.6667, P 0.6250, Lp 0.41672θ 120° — L 0.6667, P 0.6250, Lp 0.41672θ 125° — L 0.6881, P 0.6645, Lp 0.45732θ 125° — L 0.6881, P 0.6645, Lp 0.45732θ 130° — L 0.7202, P 0.7066, Lp 0.50892θ 130° — L 0.7202, P 0.7066, Lp 0.50892θ 135° — L 0.7654, P 0.7500, Lp 0.57402θ 135° — L 0.7654, P 0.7500, Lp 0.57402θ 140° — L 0.8278, P 0.7934, Lp 0.65682θ 140° — L 0.8278, P 0.7934, Lp 0.65682θ 145° — L 0.9140, P 0.8355, Lp 0.76372θ 145° — L 0.9140, P 0.8355, Lp 0.76372θ 150° — L 1.0353, P 0.8750, Lp 0.90592θ 150° — L 1.0353, P 0.8750, Lp 0.90592θ 155° — L 1.2118, P 0.9107, Lp 1.10362θ 155° — L 1.2118, P 0.9107, Lp 1.10362θ 160° — L 1.4845, P 0.9415, Lp 1.39762θ 160° — L 1.4845, P 0.9415, Lp 1.39762θ 165° — L 1.9485, P 0.9665, Lp 1.88332θ 165° — L 1.9485, P 0.9665, Lp 1.88332θ 170° — L 2.8904, P 0.9849, Lp 2.84682θ 170° — L 2.8904, P 0.9849, Lp 2.84682θ 175° — L 5.7423, P 0.9962, Lp 5.72052θ 175° — L 5.7423, P 0.9962, Lp 5.72052θ / °factor (log)

Hover anywhere on the plot to read L, P and their product at that angle — the chart is on a logarithmic scale precisely because the numbers span two orders of magnitude, and a log scale is not something to read a value off by eye. Three curves, because a linear axis would put everything past 40° on the baseline. The Lorentz factor falls steeply and then turns back up; the polarisation factor is a shallow dip bottoming at 2θ = 90°; their product has its minimum at the marked angle.

The turn is the part worth remembering. Lp does not simply decrease with angle — it reaches a minimum near 2θ = 98° and climbs again, so back-reflection lines are boosted by the same mechanism that boosts the low-angle ones. A reader carrying away only “divide by Lp, it gets smaller” has the second half backwards.

The monochromator

Both forms of P are one expression, (1 + K cos² 2θ) / (1 + K), with K = 1 for an unmonochromated beam and K = cos² 2θM for an ideally mosaic monochromator. Writing them as two formulae is how a page comes to apply one and claim the other, so there is no second formula here and no branch.

With no monochromator K = 1 and P is the familiar (1 + cos² 2θ)/2. Choose one above to see what leaving it out of the correction would cost.

cos² is the mosaic assumption and it is not universal. A perfect crystal — germanium, silicon — polarises as |cos 2θM| instead, which is larger, so a real monochromator sits somewhere between the two. Graphite is offered above because graphite really is mosaic; anything else is entered as an angle, with this assumption still in force.

The chain, run backwards

rutile at 1.541838 Å. The I column is what the pattern looks like — the simulator’s own intensities, each scaled to 100 on the strongest line. Divide by Lp and what is left is the m|F column, scaled to 100 on its strongest line.

hkl 2θ / ° d / Å m I Lp m|F mono. residual
110 27.456 3.2485 4 100.00 4.0839 37.73 +0.00 %
101 36.107 2.4876 8 45.72 2.2625 31.13 +0.00 %
200 39.220 2.2971 4 6.88 1.8852 5.62 +0.00 %
111 41.271 2.1875 8 18.93 1.6830 17.33 +0.00 %
210 44.077 2.0545 8 6.90 1.4521 7.32 +0.00 %
211 54.363 1.6876 16 58.54 0.9020 100.00 +0.00 %
220 56.670 1.6243 4 17.45 0.8207 32.76 +0.00 %
002 62.810 1.4795 2 8.40 0.6520 19.85 +0.00 %
310 64.099 1.4528 8 8.47 0.6237 20.92 +0.00 %
221 65.563 1.4238 8 0.66 0.5940 1.70 +0.00 %
301 69.061 1.3600 8 21.54 0.5325 62.30 +0.00 %
112 69.861 1.3464 8 10.55 0.5202 31.25 +0.00 %
311 72.479 1.3041 16 1.17 0.4837 3.74 +0.00 %
320 74.463 1.2742 8 0.25 0.4596 0.82 +0.00 %
202 76.605 1.2438 8 2.30 0.4369 8.10 +0.00 %
212 79.901 1.2006 16 1.32 0.4076 5.00 +0.00 %
321 82.409 1.1703 16 4.65 0.3895 18.37 +0.00 %
400 84.323 1.1485 4 3.20 0.3780 13.03 +0.00 %
410 87.559 1.1142 8 1.17 0.3623 4.98 +0.00 %
222 89.633 1.0937 8 7.95 0.3547 34.52 +0.00 %

The two columns rank the lines differently, and that is the whole point of the correction. The strongest peak and the strongest reflection are not the same line here. The tallest peak is 110; the largest ∑m|F|² belongs to 211. 222 moves 6 places — 10th strongest on the chart, 4th once Lp is off. A peak is what a detector recorded at one angle; a reflection is a point of the reciprocal lattice with a structure factor. Lp is most of what separates them, and it is why a pattern cannot be read as a table of |F|² by eye.

The last column is what it would cost to correct data from the monochromator setting above with the unmonochromated formula, measured relative to the strongest line, because that is where a reader’s own scale comes from.

The window is the simulator’s own, 5–90°, so the table stops before the Lp minimum the chart above marks. The chart is drawn over the full range; the table is not.

Absorption, and why powder people ignore it

For a flat plate in symmetric reflection — Bragg–Brentano, which is what almost every laboratory diffractometer is — the beam in and the beam out travel equal paths, and the absorption factor comes out in closed form:

A(θ) = [1 − exp(−2μt / sin θ)] / 2μ

At large μt that is 1 / 2μ, with no θ left in it. Absorption does not drop out of powder work because it is small — a millimetre of rutile stops the beam in the first ten micrometres — it drops out because in this one geometry it is the same for every line, so it joins s in the scale factor nobody needs to know. Make the sample thin and it stops being constant, which is why a thin film or a smear mount on a zero-background wafer needs the full expression.

A capillary is the case with no closed form at all: the path length depends on where in the cylinder the scattering happened, and the integral has no elementary answer. That is why International Tables volume C tabulates A* against μR instead of giving an equation, and it is why this page does not compute it rather than fitting a polynomial nobody could check.

rutile at Cu Kα: μ/ρ = 129.27 cm²/g on a density of 4.247 g/cm³, so μ = 549 cm−1. At 1 mm thick:

2θ / ° depth holding 90 % / µm depth holding 99 % / µm A(θ) ÷ 1/2μ
10 1.8 3.7 1.000000
30 5.4 10.9 1.000000
60 10.5 21.0 1.000000
90 14.8 29.7 1.000000
140 19.7 39.4 1.000000

The last column is the ratio to the infinitely-thick value, and it is 1 at every angle for any sample thick enough — that is the constancy claim, computed rather than asserted. Set the thickness above to something genuinely thin and watch it stop being 1, first at low angle, where the beam travels furthest for a given depth.

The depths run the other way from the intuition. They carry sin θ, so the low-angle lines are the shallow ones: the first peak in a pattern is sampling a thinner slab of the specimen than the last one. On a sample with a gradient — a surface layer, a preferred-orientation skin from pressing the powder — that is a systematic difference between lines and not a scale factor.

What this page leaves out

Named rather than silently absent, because a correction nobody mentions is a correction nobody applies:

  • Preferred orientation. Not a correction so much as a defect in the specimen: the crystallites are not randomly oriented, so m stops describing how many of them are in the diffracting condition. It is modelled (March–Dollase) rather than computed, and it needs a parameter refined against the data.
  • Extinction. A strong reflection can be attenuated by re-diffraction inside a single mosaic block, which makes the strongest lines too weak — a correction that depends on |F|, so the thing being measured appears in its own correction.
  • Thermal diffuse scattering. It sits under the peaks and rises with angle, so it is absorbed into the background and then into the displacement parameters.
  • The capillary absorption factor A*, for the reason given above.

Everything on this page is exact arithmetic on stated assumptions. Everything in this list needs either a model or a measurement, which is a different kind of claim.

Where this sits

Upstream: peak finding turns a scan into positions, heights and areas — the integrated area is the quantity this page corrects, not the peak height. Downstream: the structure factor is what |F|² is compared against, and the Wilson plot is the first thing done with a corrected data set.

Beside it: absorption computes μ/ρ for any formula, which is where the coefficient in the panel above comes from.