Crystal Centring Simulator
https://xraytools.com/centring
Before a single-crystal measurement the crystal has to sit where every axis of the goniometer meets, or it wanders out of the beam as the axes turn. Centring is done by eye through the video microscope, with the two sledges and the height screw of the goniometer head. This page lets you practise it on a crystal that starts in a different place every time.
- You supply
- Nothing of your own. Choose a diffractometer, a mount and whether the crosshair can be trusted; then turn φ and the three screws until the crystal stays put.
- Reading it
- The picture shows two directions of the crystal’s offset; the third, depth, shows only as blur. One view is therefore never enough, and the final check is a full turn of φ. The sledges turn with φ, so the screw that moves the crystal sideways changes as you rotate.
See also: Absorption Coefficient Calculator · The Ewald Construction
Centre the crystal
One turn moves a sledge 0.5 mm: − and + are a hundredth of a turn (5 μm), −⅒ and +⅒ a tenth (50 μm). Hold a button to keep turning.
Exercise 1. The crystal starts away from the centre.
The picture is the video microscope, 1.2 mm across. The red crosshair should mark the rotation centre; the dashed circle is the size of the beam. A crystal well behind or in front of the centre is out of focus; within the depth of field, ±80 μm here, it looks sharp wherever it is.
The head diagram shows the goniometer head from above. The two sledges, 1 and 2, turn with φ; each arrow shows which way a + turn moves the crystal. A faded arrow points at the camera: that sledge moves the crystal along the view, which shows only as blur and not at all within the depth of field, and its screw is marked “along the view”. The height screw moves the crystal along the pin.
How centring is done
Every axis of the goniometer passes through one point, the instrument centre, and the beam passes through it too. A crystal on that point stays in the beam whatever the axes do. The goniometer head has three screws to put it there: two sledges at right angles that move the crystal across the φ axis, and a height screw that moves it along the pin.
- Go to the centring position. On a fixed-χ goniometer the φ axis is tilted by χ = 54.74° and cannot be made upright, so it crosses the picture at that angle. The camera then looks at right angles to φ.
- Set the height first. Turn the height screw until the crystal sits on the crosshair along the line of the pin. Height does not change when φ turns, so it is done once.
- Turn φ until one sledge points across the picture. That sledge moves the crystal sideways on screen. The other one now points at the camera and moves the crystal along the view, which the picture shows only as blur, and not at all for a small error.
- Bring the crystal onto the axis with the sideways sledge. Then turn φ by 180°. If the crystal jumps to the other side, it was not on the axis: move it back half the jump. Halving needs no trust in the crosshair, because the axis lies midway between the two positions.
- Turn φ by 90° so the other sledge points across, and repeat steps 3 and 4 with it.
- Check. Spin φ through a full turn: the crystal should stay in the beam circle and in focus. Then turn ω. At the centring position an error towards or away from the camera does not move the crystal on screen: a large one blurs it, and one inside the microscope’s depth of field is invisible from this view. That is why a second φ angle is needed, and why any other orientation of the axes turns the same error into a visible shift.
Check the centring gives a percentage: the crystal’s distance from the instrument centre measured against the beam radius, 60 μm, so 100 % is on the centre, 0 % a beam radius or more away, and 75 % (15 μm) is the least it accepts as centred.
Centre the crystal, not the loop or the drop of oil around it. A loop seen edge-on is a thin line; seen face-on its film can hide the crystal’s edges.
The instrument here is perfect: its axes meet exactly. Real goniometers are calibrated so that the crosshair marks the centre; when it does not, the 180° halving still works. The picture is a schematic of the geometry, drawn with ω vertical for both instruments; a real instrument places its axes and camera in its own way. Fixed-χ three-circle goniometers include the Bruker SMART APEX platform; κ goniometers include those of Oxford Diffraction instruments such as the SuperNova.